How do Electron Microscopes Work? ๐Ÿ”ฌ๐Ÿ› ๐Ÿ”ฌ Taking Pictures of Atoms

Branch Education

19m 54s

Introduction to Electron Microscopy

  • Scientists/engineers design transistors width of DNA strand
  • Electron microscopes changed how humanity sees micro/nanoscopic world
  • European peacock butterfly wing example: light microscope shows overlapping scales, electron microscope reveals detailed scale texture
  • Used for 60 years to develop smaller transistors
  • Modern microscopes zoom millions of times, capture individual atom images

Types of Electron Microscopes

  • Scanning Electron Microscope (SEM)
    • Used for surface images (butterfly wing, toothbrush bristles, body cells with bacteria)
    • Maximum resolution: 1 nanometer
    • Resolution = spacing between two adjacent resolvable features
  • Transmission Electron Microscope (TEM)
    • Images structures inside materials (like X-ray for bones)
    • Takes pictures of transistor sections, proteins in mitochondria, gold nanoparticles
    • More complex than SEMs
    • Resolution: up to 50 picometers (hydrogen atom size)

TEM Basic Operation Principle

  • Generates electrons, accelerates to 70% speed of light
  • Magnetic lenses focus electrons to small area
  • Electrons transmitted through specimen
  • Different densities/materials scatter electrons differently
  • Imprints image onto electron beam
  • Objective lens magnifies 40 times
  • Projector lenses magnify additional 50,000 times
  • Final image: 5cm wide, captured by high-resolution camera

Light vs Electron Limitations

  • Visible Light Limitations
    • Physically limited to 2,000x magnification
    • Further zoom results in blur without detail
    • Average wavelength: 540nm
    • Cannot resolve features <300nm apart due to diffraction
  • Electron Advantages
    • Meaningful magnifications up to 2 million times
    • Wavelength: 2.5 picometers (200,000x smaller than visible light)
    • Theoretical resolution: 1 picometer
    • Actual resolution: 50 picometers (due to magnetic lens limitations)

Scale Comparison

  • Micrometers โ†’ nanometers โ†’ picometers size progression shown

Field Emission Source (Electron Generation)

  • Tungsten Crystal Needle
    • Few nanometers wide
    • Negatively charged electrons attracted to positive electric fields
  • Extractor Ring
    • Connected to positive 5,000 volts
    • Pulls electrons from tungsten
    • Sharp point amplifies electric field effect
    • Frees electrons from tungsten

Electron Acceleration System

  • Series of metal rings with tens of thousands of volts between them
  • Positively charged rings attract negatively charged electrons
  • Accelerates electrons to 70% speed of light
  • Two key reasons for high speed:
    • Travel through specimen (typically 100nm thick)
    • Shorter wavelength = higher resolution

Vacuum System

  • Vacuum pumps remove atmospheric molecules
  • Creates vacuum similar to outer space
  • Prevents electron scattering from air molecule collisions
  • Essential for image quality

Magnetic Lens Systems

  • Three main sets: condenser, objective, projector
  • Condenser Lenses
    • Focus electrons from source
    • Project onto sample
    • Illuminate area: micrometer to several nanometers
  • Apertures/Holes
    • Filter electrons fanning from center
    • Create more parallel electron beam

Specimen Handling

  • Specimen Holder
    • Inserted through air lock into vacuum chamber
    • Moves in all three directions (X, Y, Z)
    • Rotates along X-axis, some also Y-axis
    • Enables perpendicular imaging of features (transistors)

Image Formation Process

  • Electron beam hits specimen with different elements/densities
  • Electrons scatter differently
  • Imprints image on transmitted electron beam
  • Objective Lens (heart of microscope)
    • Magnifies 40 times
    • Optical aberrations define final resolution
  • Projector Lenses
    • Magnify additional 50,000 times
    • Final image width: few centimeters

Optical Aberrations and Resolution Limits

  • Current limitation: not electrons, but lens aberrations/distortions
  • Main types: spherical, chromatic aberrations
  • Problem: perfectly controlling electron beam extremely difficult
  • Effect: aberrations add blurriness, impede resolution
  • Cascading effect: projector lens magnifies objective lens aberrations plus adds own
  • Focus: considerable science/engineering dedicated to reducing objective lens aberrations

Magnetic Lens Operation

  • Construction
    • Copper wire coil surrounded by iron housing
    • Current through coils produces magnetic field
    • Iron routes field to pole pieces
    • Channeled into optical column
  • Physics: Lorentz Law
    • Force = charge ร— velocity ร— magnetic field
    • Electrons with velocity away from optical axis forced toward center
    • Electrons traveling down center experience no lens force
    • Acts as convex/converging lens
  • Result
    • Focuses electrons to focal point
    • Past focal point: electrons expand, produce magnified image
    • Magnification depends on: magnetic field strength, lens position, detector/camera position

Image Detection Systems

  • System 1: Phosphorescent Screen
    • Special coating glows when electrons hit
    • Camera views screen
    • Used for microscope alignment and specimen overview
  • System 2: CMOS Camera
    • Higher resolution and dynamic range
    • More sensitive
    • Used for high-resolution image capture
    • Phosphorescent screen moves out of way
  • Purpose of dual system: protect sensitive CMOS camera from incorrectly focused beam damage

Additional Equipment/Features

  • X-ray detectors, energy filters, phase plates
  • Monochromators, multiple correctors
  • Aperture adjustment mechanisms
  • Water cooling for magnetic lenses
  • Circuitry for magnetic lens/field emission source control
  • Vacuum pumps, power supplies
  • Air cushions for vibration isolation

Scanning Electron Microscope (SEM) Details

  • Operation Principle
    • Creates focused spot instead of illuminating entire area
    • Scans spot across object
    • Electrons bounce off, create secondary electrons, back-scattered electrons, X-rays
    • Measures signals for surface topology and chemical composition
  • Applications: butterfly wing, salt crystal images
  • Limitations
    • Surface-only imaging
    • Resolution limited by focused spot size and electron penetration depth (interaction volume)
    • Practical resolution: ~1 nanometer

Scanning Transmission Electron Microscope (STEM)

  • Hybrid approach: similar to TEM but focuses beam into spot like SEM
  • Operation: uses deflection coils to scan spot through specimen
  • Benefits
    • Different image contrast mechanism
    • Paired with X-ray detector enables elemental analysis
  • Versatility: expensive TEMs typically have optical elements for both TEM and STEM modes

Sponsor Information

  • Thermo Fisher Scientific
    • Provided 3D model and technical assistance
    • Manufactures wide range of laboratory equipment
    • Centrifuges, incubators, X-ray/mass spectrometers, PCR systems (COVID-19 testing)
    • Focus on supporting scientific research and innovation

TwinMind

Get summaries like this for your own meetings

TwinMind takes notes in the background during meetings, lectures, and calls, then writes the summary for you.

Free plan to get started

No bot joins your calls

Phone, desktop & Chrome apps

Your next meeting, summarized like this

TwinMind listens in the background and turns meetings, lectures, and calls into searchable notes you can chat with. Start free.

Available on phone, desktop & Chrome

Your next meeting, summarized like this

TwinMind listens in the background and turns meetings, lectures, and calls into searchable notes you can chat with. Start free.

Available on phone, desktop & Chrome

Your next meeting, summarized like this

TwinMind listens in the background and turns meetings, lectures, and calls into searchable notes you can chat with. Start free.

Available on phone, desktop & Chrome